Product Comparison
Duke University Medical Center Research
In 2005, an independent study was done to determine how the performance of ALBAclone? blood grouping reagents compared with the performance of marketed products. The findings of this research are reported below.
Download the product comparison form.
| Test Reagent | Reference Reagent 1 | Sample Size | Duke Results | Download Details |
|---|---|---|---|---|
| ALBAclone® Anti-A (Z001U) | Ortho/MTS Anti-A* | 505 | No typing discrepancies were identified | |
| ALBAclone® Anti-B (Z015U) | Ortho/MTS Anti-B* | 505 | No typing discrepancies were identified | |
| ALBAclone® Anti A,B (Z023U) | Ortho/MTS Anti-A, Anti-B* | 505 | No typing discrepancies were identified | |
| ALBAclone® Anti-D alpha (Z031U) | Ortho/MTS Anti-D* | 405 | No typing discrepancies were identified with ALBAclone® products** | |
| ALBAclone® Anti-D beta (Z036U) | Ortho/MTS Anti-D* | 405 | No typing discrepancies were identified with ALBAclone® products** | |
| ALBAclone® Anti-D delta (Z039U) | Ortho/MTS Anti-D* | 405 | No typing discrepancies were identified with ALBAclone® products** | |
| ALBAclone® Anti-D blend (Z041U) | Ortho/MTS Anti-D | 405 | No typing discrepancies were identified with ALBAclone® products** | |
| ALBAclone® Anti-c blend (Z083U) | Ortho/MTS Anti-c | 200 | No typing discrepancies were identified | |
| ALBAclone® Anti-Lub (Z223U) | Anti-Lub patient sample | 100 | No typing discrepancies were identified† | |
| ALBAclone® Anti-E (Z073U) | Ortho Anti-E | 200 | No typing discrepancies were identified | |
| ALBAclone® Anti-M (Z171U) | Gammaclone® Anti-M | 100 | No typing discrepancies were identified | |
| ALBAclone® Anti-N (Z176U) | Immucor Anti-N | 100 | No typing discrepancies were identified | |
| ALBAclone® Anti-k (Z137U) | Ortho Anti-k | 100 | No typing discrepancies were identified | |
| ALBAclone® Anti-Lea (Z212U) | Ortho Anti-Lea | 100 | No typing discrepancies were identified with ALBAclone® products†† |
|
| ALBAclone® Anti-Leb (Z217U) | Ortho Anti-Leb | 100 | No typing discrepancies were identified with ALBAclone® products‡ |
* Reagents were from Lot 403715. For ALBAclone Anti-A (Z001), another reagent was also tested: Immucor Anti-A Lot #101640.
** In one sample with mixed field agglutination, the Ortho reference agent produced equivocal results, while the ALBAclone® product produced unequivocal results.
† When the test results were compared with the performance of the comparator reagent, no typing discrepancies were identified although no result was recorded for the reference reagent with sample 1007. The reagent performed as specified and no unexpected reactions were observed.
†† When the test results were compared with the comparator reagent, one test was repeated as the reference failed to react originally. The results were comparable on repeat. The reagent performed as specified and no unexpected reactions were observed.
‡ One test was repeated as the test failed to react originally. The results were comparable on repeat. Another sample gave a discrepant result; however, the result obtained with the ALBAclone® antisera was confirmed with Immucor® Anti-Leb.
For quick answers to many product questions, take a look at our Technical FAQs.